Measurements of PEEP-induced changes in lung volume - Evaluation of a laser monitor

Uhlig, T., Kondo, T. and Sly, P. D. (1997) Measurements of PEEP-induced changes in lung volume - Evaluation of a laser monitor. Chest, 112 1: 107-112. doi:10.1378/chest.112.1.107

Author Uhlig, T.
Kondo, T.
Sly, P. D.
Title Measurements of PEEP-induced changes in lung volume - Evaluation of a laser monitor
Journal name Chest   Check publisher's open access policy
ISSN 0012-3692
Publication date 1997-07
Sub-type Article (original research)
DOI 10.1378/chest.112.1.107
Volume 112
Issue 1
Start page 107
End page 112
Total pages 6
Place of publication Northbrook, IL, United States
Publisher American College of Chest Physicians
Language eng
Formatted abstract
Study objective: To evaluate a newly developed laser monitor in the measurement of positive end-expiratory pressure (PEEP)-induced changes in end-expiratory lung volume (EELV) in spontaneously breathing subjects.

Design: An open comparison between two different methods to assess breathing parameters.

Setting: A respiratory research unit.

Participants: Six spontaneously breathing, healthy volunteers.

Interventions: Stepwise increases of PEEP from 0 to 0.8 to 1.6 kPa during spontaneous breathing; repeated validation of the laser monitor in each subject.

Measurements and results: Pressure and flow were recorded at the airway opening. Abdominal wall displacement (AWD) measured by the laser sensor was recorded simultaneously. The time lag between the volume and the laser signals during baseline was 0.068 ± 0.052 s and during maximal PEEP, 0.1084 ± 0.093 s. There was no baseline drift in either the PEEP or the AWD signals. Mean EELV decreased by 290 mL to 1,157 mL after the PEEP valve was removed. Within each individual, the ratio between EELV and AWD showed only small variations. Measurements of tidal volume (V(T)) and AWD showed good agreement at all PEEP levels. Mean V(T) decreased in all but one subject during PEEP. With the increase in PEEP, the end-expiratory abdominal baseline increased linearly over a large range of end-expiratory pressures with a flattening of the curve at high PEEP levels in all subjects.

Conclusions: The laser monitor is sufficiently accurate for measuring PEEP-induced changes in EELV during spontaneous breathing in healthy adult subjects. Monitors incorporating multiple laser sensors may have considerable clinical promise.
Keyword End-expiratory lung volume
Laser monitor
Lung function measurements
Q-Index Code C1
Q-Index Status Provisional Code
Institutional Status Non-UQ

Document type: Journal Article
Sub-type: Article (original research)
Collection: Faculty of Health and Behavioural Sciences -- Publications
Version Filter Type
Citation counts: TR Web of Science Citation Count  Cited 4 times in Thomson Reuters Web of Science Article | Citations
Scopus Citation Count Cited 0 times in Scopus Article
Google Scholar Search Google Scholar
Created: Wed, 17 Nov 2010, 11:56:09 EST